Setup – Median Sensory NCS

0
6120
14-cm distance (black to black)

Electrode placement (use the ring electrodes)

14-cm distance (black to black)
  • E1 (black):  Proximal phalanx of the index finger — want to keep away from the metacarpophalangeal (MCP) joint skin, as this may pick up motor conduction artifact.
  • E2 (red):  3-4cm distally from E1.
  • Ground:  Dorsum of hand

Stimulation site

  • (-) Cathode (black):  14-cm proximal to E1.  Note: Make the 14-cm line to the middle of the wrist crease, then proximal along the median nerve.
  • (+) Anode (red):  proximal (in line with the median nerve)

Additional Information

  • Type of stimulation = Antidromic
  • May need to place the recording electrode (E1) more distally if you are getting too much motor volume-conduction
  • CPT codes for nerve conduction studies
    • 95900 Nerve conduction, amplitude and latency/velocity study, each nerve; motor, without F-wave study
    • 95903 Nerve conduction, amplitude and latency/velocity study, each nerve; motor with F-wave study
    • 95904 Nerve conduction, amplitude and latency/velocity study, each nerve; motor with F-wave study, each nerve; sensory
    • 95934 H-reflex, amplitude and latency study; record gastrocnemius/soleus muscle
    • 95936 Record muscle other than gastrocnemius/soleus muscle

LEAVE A REPLY

Please enter your comment!
Please enter your name here